To comply to AAMI standard on Endotoxins( <0.25IU /ML)-this was not being achieved.
To prevent Infection control in patients undergoing haemo-dialysis.
To introduce best practices which are currently not being followed (For example the customer was using formaldehyde for sanitation which remained in the loop resulting in improper disinfection).
Client: NH, Guwahati Location: India
Metcon’s Solution
Metcon’s intricate understanding of the water application for Dialysis led to innovation and
implementation of a fool-proof soln.
- Design of heat sanitation distribution loop(80 deg.C) after 2 pass RO - Introduction of PAA & H2O2 for cold disinfection.
Business Benefits
Reliability of hospital’s dialysis facility due to strict control over quality wrt endotoxins.
Consistent treated water quality (patients need not fear of infections & Raigad effect normally associated with water which do not meet stringent AAMI standards).
Key Challenges:
The hospital governing body wanted to serve safe drinking water to all his staffs, patients & their relatives as well as maintain a regulation about water wastage by mishandling - wanted to adopt innovative systems for distribute safe drinking water without engage any manhead for operation as well as wastage restrict.
Metcon was called in to design and implement an innovative solution that is not usual for the industry.
Client: NH, Guwahati Location: India
Metcon’s Solution
After successfully designing the 150 LPH Automatic Water Treatment Machine (Water ATM).
The system is running since April’ 2018.
Business Benefits
The customer continued to reap the benefits of their Customer and Employee satisfaction & enjoy
leadership role in their industry segment.
Operational efficiency achieved through saving water (through restrict wastage of water).